Measuring range
MR I: 150 ... 800 °C
MR II: 180 ... 1200 °C
MR II: 250 ...2000°C
Sensors
photo diode
Spectral sensitivity
1.8 - 2.2 µm
Focussing range MR I
0.3 m ... ∞ (standard lens)
Focusing range MR II
0.2 m - 0.4 m (close-up lens)
0.4 m - ∞ (standard lens)
1.2 m - ∞ (telephoto lens)
Focusing range MR III
0.2 m - 0.4 m (close-up lens)
0.4 m - ∞ (standard lens)
0.2 m - ∞ (wide-angle lens)
1.2 m - ∞ (telephoto lens)
Distance to target-size ratio MR 1
40 : 1 at 400 mm (standard)
Distance to target-size ratio MR II
56 : 1 at 400 mm (close-up)
60: 1 at 400 mm (standard)
96 : 1 at 1200 mm
(telephoto)
Distance to target-size ratio MR III
200 : 1 at 400 mm (close-up)
210 : 1 at 400 mm (standard)
310 : 1 at 1200 mm
(telephoto)
55 : 1 at 400 mm (wide angle)
Digital output
Periodic output of measure-ment data with adjustable cycle time
Analogue output 1 & 2
0(4) ... 20 mA linear,
switchable, scalable, output insulated from power supply
(4…20 mA normally)
Resistance
max. 500 Ω
Response time t98 MR I
≤ 50 ms (T > 150 °C)
≤ 15 ms (T > 200 °C)
≤ 2 ms (T > 350 °C)
Response time t98 MR II
≤ 75 ms (T > 180 °C)
≤ 35 ms (T > 200 °C)
≤ 5 ms (T > 300 °C)
≤ 2 ms (T > 600 °C)
Response time t98 MR III
≤ 50 ms (T > 250 °C)
≤ 2 ms (T > 750 °C)
Resolution
Analogue output
0.2 K + 0.03 %
of the adjusted span
Display
1 K
USB / RS 485
0.1 K at terminal operation
Measuring uncertainty
MR I and II
0.75 % of range and but
at least 5 K
(at ε =1.0 and TA=23 °C)
Measuring uncertainty
MR III
0.5 % of range and but
at least 4 K
(at ε =1.0 and TA=23 °C)
Linearisation
digital via microcontroller
Repeatability
1 K
Sighting device
through-the-lens sighting with target marking, laser spot light or camera
Ambient operating
temperature
0 ... 65 °C
Indication of Overheating
If the internal temperature of the pyrometer exceeds 80 °C the analogue output will show a value > 20.5 mA.
Storage temperature
-20 ... 80 °C
Temperature coefficient with reference to 23 °C
0.25 K / K (for T < 500 °C)
0.05 % / K (for T ≥ 200 °C)
of temperature reading
Data communication
USB / RS485 with integrated software to set parameters and transmit measurement data to a PC
Analogue input
0 - 10 V
Digital output
2 Open collector outputs
24 V; ≤ 30 mA
Digital input
2 to 24 V
Power supply requirements
24 V DC +10% / -20%
current input ≤135 mA
150 mA with switched on spotlight
Ripple: ≤ 200 mV
Dimensions
Ø 65 x 220 mm
(when connector is attached)
Housing material
Stainless steel
Mounting
External thread M 65 x 2
length 40 mm
Weight
approx. 0.9 kg
Connection
with 8-pin connector
Protection
IP 65 according to DIN 40050
(when connector is attached)
Adjustable parameters
Analogue output 1 & 2
source/ scaling
Digtal input output 1 & 2
source/ switch point
Transmission factor
Compensation of background radiation
Look-up table for temperature alignment
Emissivity ε
10.0 bis 99 %
increment size 0,1 %
Smoothing function t98
0 - 16 sec
Memory modes
- Min./Max. (peak picker)
- Double maximum with
adjustable hold time
Optional accessories
Calibration certificate
according to ISO 9001
Calibration certificate
according to DKD
Large variety of mounting
devices, digital displays,
software, etc.